Cleaning and decontamination
Predisinfection:
1. Remove all samples from the sample chamber and store them properly.
2. Remove accessories from the biological safety cabinet and disinfect them using
the disinfection procedure recommended by the manufacturer.
3. The workplate and stainless steel components can be removed from the sam-
ple chamber and disinfected separately.
4. For predisinfection, spray disinfectant on all sample chamber surfaces or wipe
the surfaces using disinfectant.
5. Do not remove the optional UV lamps from the sockets; wipe them thorou-
ghly using a damp cloth.
6. Switch the device to work mode, move the front window to the work
position.
7. Allow disinfectant to react as recommended by the manufacturer, then operate
the biological safety cabinet for at least 15 to 20 minutes in the work mode so
that released aerosols can be absorbed by the filters.
Cleaning:
1. Remove dirt residues and deposits thoroughly using a solution of tepid water
and commercially available dishwashing agent.
2. Wipe the surfaces clean using a clean cloth and plenty of clear water.
3. Remove the cleaning liquid from the floorpan and wipe all sample chamber
surfaces dry.
Final disinfection:
1. Again, spray disinfectant on all sample chamber surfaces or wipe the sur-
faces clean using disinfectant.
2. Allow disinfectant to react as recommended by the manufacturer.