Cleaning and decontamination
9.3
UV disinfection after a wipe/spray disinfection
A UV disinfection can be performed either by using the optional integral UV lamps
or by using a mobile UV device.
9.3.1
UV disinfection using the integral UV lamp
To start the UV disinfection routine:
1. Close the front window completely, the air system operates in the reduced
mode (yellow LED illuminated).
2. Start UV disinfection using the
UV
key at the operating panel:
Keep the
UV
depressed until the display alternatingly shows dIS and the re-
maining disinfection time.
To interrupt / cancel the UV disinfection routine:
1. Press the
UV
key (the display shows the operating hours).
2. Slide the front window up.
9.3.2
UV disinfection using a mobile UV device (optional)
To control the disinfection routine of a mobile UV device with the biological safety
cabinet software, ensure that the connecting plug of the UV device is compatible
with the UV disinfection adapter of the biological safety cabinet.
To start the UV disinfection routine:
1. Position the mobile UV device centered on the work area and connect it to
the UV disinfection adapter.
2. Close the front window completely, the air system operates in the reduced
mode (yellow LED illuminated).
3. Start UV disinfection using the
UV
key at the operating panel:
Keep the
UV
depressed until the display alternatingly shows dIS and the re-
maining disinfection time.
To interrupt / cancel the UV disinfection routine:
1. Press the
UV
key (the display shows the operating hours).
2. Slide the front window up.
9.3.3
Changing UV disinfection time
The UV disinfection time is preset; this setting can be changed, if required:
1. Switch the cabinet on and move the front window to the work position.
2. Keep the UV key depressed until the preset time is displayed.
3. To increase the time:
Press the
I
NTERNAL POWER SUPPLY
key. Each pressing of the key increases the
time by 30 minutes.
4. To reduce the time:
Press the
I
NTEGRAL SAMPLE CHAMBER ILLUMINATION
key. Each pressing of the key
reduces the time by 30 minutes.
5. To save the values:
Press the
UV
key.
The display shows the operating hours again.