When a manual cleaning is necessary, use the following procedure:
1.
On the home screen, press Settings
4
Clean Instrument.
The instrument door unlocks allowing access to the consumables.
2.
Remove the Ion S5
™
Wash Solution bottle to access the waste reservoir, then
remove and empty the waste reservoir.
1
2
3
3.
Reinstall the empty waste reservoir and a used Ion S5
™
Wash Solution bottle.
4.
Ensure that the Ion S5
™
Sequencing Reagents cartridge and Ion S5
™
Wash
Solution bottle are properly installed.
IMPORTANT!
Perform the cleaning with a used reagent cartridge and wash
solution bottle installed. The cleaning procedure pumps cleaning solution into
the wash solution bottle and reagent cartridge making them unsuitable for
sequencing.
5.
Place a used sequencing chip in the chip clamp, then push the chip clamp in all
the way to engage.
6.
Close the instrument door, then press Next.
Cleaning takes ~35 minutes to complete. On completion, the instrument door
automatically unlocks and the chip and cartridge clamps disengage.
Chapter 3
Instrument operation
Maintain the Ion GeneStudio
™
S5 Food Protection Instrument
3
Ion GeneStudio
™
S5 Food Protection Instrument User Guide
15