Instrument installation by trained personnel only
IMPORTANT!
The Ion GeneStudio
™
S5 Food Protection System is installed by trained
service personnel and must not be relocated without assistance from trained service
personnel. See “Food Safety support“ on page 53.
Nucleic acid contamination
IMPORTANT!
A primary source of contamination is DNA fragments from previously
processed samples. Do not introduce amplified DNA into the library preparation
laboratory or work area.
Instrument vibration and clearances
IMPORTANT!
Significant vibration during sequencing may add noise and reduce the
quality of the measurements. The Ion GeneStudio
™
S5 Food Protection System must
be installed on a bench that is free from vibrations or in contact with equipment that
can cause vibrations to the bench (freezers, pumps, and other similar equipment).
IMPORTANT!
Place the instrument at least 40 in. (1 meter) away from major sources
of electronic noise such as refrigerators or microwaves.
Chapter 1
Product information
Instrument installation by trained personnel only
1
Ion GeneStudio
™
S5 Food Protection Instrument User Guide
7