Item
Description
When/How to use
Pulse clean
Provides step by step instruction to perform a pulse clean
of the instrument if a solid
‑
state buffer leak occurs.
Use if directed to do so by Technical
Support as part of a troubleshooting
procedure:
1. Install consumables and a used
sequencing chip.
Note:
Install Cleaning solution in
both the Wash solution (1) and
Cleaning solution (2) locations.
Use an empty Reagent cartridge.
2. Close the instrument door, then
press Next.
The pulse cleaning procedure starts
automatically.
Upload
Diagnostics
Allows upload of instrument diagnostics files for review by
support personnel.
For troubleshooting if directed to do so
by Technical Support.
Note:
The system must be connected
to an active network port. If no network
is present, the file is pushed to the
Torrent Browser. You can then
download the file and email it to
Technical Support.
To retrieve the file using the Torrent
Browser, navigate to Help
4
Instrument
Diagnostics, then follow the on-screen
instructions.
Appendix A
Touchscreen reference
Settings
A
Ion GeneStudio
™
S5 Food Protection Instrument User Guide
25