S.U.B. TK Resistive Quick Start Guide
Quick Start Guide
11. Ensure the cable has been routed through the cable and tubing
organizer and connect to the fitting located at the rear of the
control cabinet labeled BAG PRESS (
●
● ●
).
12. IMPORTANT
Disengage any clamps on the lineset for the
exhaust filter(s). If the clamps are left engaged, the BPC could
burst when filling it with air or liquid.
13. Position and close a bar clamp (Part No. SV20664.01) on the
bottom drain line as close as possible to the BPC port (Figure 11).
Also place bar clamps on probe ports intended for conventional
probes (Figure 12).
14. Insert RTD or selected temperature sensor into thermo-well
(Figure 13, Figure 14).
• Place small amount of glycerol (0.5 mL) in well to aid in heat
transfer.
• Sensor should be inserted until base of probe meets the mouth
of the thermo-well.
• Secure by twisting the luer lock collar. The thermo-well will
stretch slightly when RTD is seated.
Figure 13
Figure 11
Figure 14
Figure 3.12