Thermo Scientific
8
S.U.B. TK Resistive Quick Start Guide
Quick Start Guide
5. Guide bottom tubing lineset through their proper channels:
• Insert frit and open pipe sparger lines through the sparge plate
(round, white plastic with slots) first, before locating it in the
base of unit.
• Insert drain line through bottom front face directly below
agitation motor. Hang the drain line coil on the arm provided.
Also route the sample line through the snap in slots on the
drain coil arm.
• Attach BPC positioning tabs (bottom of chamber) onto
the alignment pins located at the base of the S.U.B. tank.
Note that the 50 L BPC does not require this step and is not
recommended, but has tab holders available, if desired.
• Insert sensors, KleenPaks, and auxiliary linesets through the
“probe belt opening on the tank”.
6. Uncover vent filters, wrap heating blankets (if used) around
the filters, attach/adjust vent bracket, and mount vent filters in
bracket.
Figure 7.
BPC tab holders used for proper BPC positioning
Figure 8.
Filter being placed in bracket