When using the optional temperature
monitoring probe, the system may be
cali4°C from the factory setting.
Begin by pressing the down arrow, the
up arrow, the Silence button, and then
the down arrow twice.
The screen below will appear on the display.
Press the Temperature button beneath Temp.
Using the up and down arrows, increase or decrease the temperature value
to match an independent, accurate temperature measuring device. When
selected, press the Time, Speed, or Temp button to save the setting.
The display will return to the Calibrate - RPM Temp screen. If nothing is
pressed for about fifteen seconds, the display will revert to the Operating
Screen and the setting will be automatically saved to memory.
Model 416/430 Orbital Shaker control system monitors and provides
alarms for six operating parameters:
Parameter . . . . . . . . . . . . . . . . . . .Alarm Message
Cycle Status . . . . . . . . . . . . . . . . .Cycle Complete
Loss of Input Power . . . . . . . . . . . . .Power Failure
RPM versus Setpoint . . . . . . . . . . .RPM Tracking
Drive Belt Integrity . . . . . . . . . . . . . . .Check Belt
Platform Movement Status . . . . . .Platform Stalled
Motor Fuse Integrity . . . . . . . . . . . . . .Check Fuse
Orbital Shakers 2-11
Thermo Scientific
Section 2
Operation
Calibrating Optional
Temperature Probe
Setting the Remote
Alarms
Summary of Contents for Forma 416
Page 1: ...Model 416 and 430 Forma Orbital Shakers Operating and Maintenance Manual 7000416 Rev 5...
Page 34: ...Orbital Shakers 2 15 Thermo Scientific...
Page 37: ......
Page 39: ......
Page 47: ......
Page 49: ......
Page 50: ...Orbital Shakers 7 1 Thermo Scientific...
Page 51: ...7 2 Orbital Shakers Thermo Scientific...
Page 54: ...Orbital Shakers 9 1 Thermo Scientific...