The electrical power requirements of the shakers are:
90-125 VAC, 50/60 Hz, 1 PH, 1.0 FLA
180-250 VAC, 50/60 Hz, 1PH, 0.5 FLA
Because these Orbital Shakers operate at a wide range of voltages, the
electrical line cord is not permanently wired to the unit. Instead, a line
cord receptacle is provided which accepts power cords for 115 or 230 volts.
A voltage selector switch is located on the rear panel of the shaker adjacent
to the line cord receptacle. Set the switch for the voltage in use at the
laboratory site, using a small flatblade screwdriver. Refer to the illustration
below, and to Figure 1-16.
Figure 1-7.
Voltage Selector Switch
Caution
Before plugging in the shaker, verify that the voltage select switch
is set for the electrical service available.
▲
The electric power switch (mains disconnect) is located on the lower front
corner of the left side panel.
Each Flask Clip up to 6.0 liters in size comes with a metal spring which
must be installed onto the clip. For flask clips through 500 ml, insert the
end of each spring into the holes on the top of the clip leg as shown in
Figures 1-8 and 1-9.
Orbital Shakers 1-5
Thermo Scientific
Section 1
Installation and Start-Up
Electrical
Requirements
Assembling the Flask
Clips
Figure 1-8.
250/300ml flask w/ cushion pad
Figure 1-9.
One screw flask clip assy
Summary of Contents for Forma 416
Page 1: ...Model 416 and 430 Forma Orbital Shakers Operating and Maintenance Manual 7000416 Rev 5...
Page 34: ...Orbital Shakers 2 15 Thermo Scientific...
Page 37: ......
Page 39: ......
Page 47: ......
Page 49: ......
Page 50: ...Orbital Shakers 7 1 Thermo Scientific...
Page 51: ...7 2 Orbital Shakers Thermo Scientific...
Page 54: ...Orbital Shakers 9 1 Thermo Scientific...