Compatible Sensors for Smart-Vue Pro Duo / Quatro Data Logger
CO
2
, Temperature and Humidity Smart Sensor | 19
The CO
2
sensor is contained inside the casing with the dual
temperature and humidity sensor attached to the back, as
shown here:
Figure 23. Sensor locations on unit
Key Specifications
•
Measured ranges
•
CO
2
:
0 to 9.99% CO
2
•
Temperature:
0°C to +50°C (+32° to +122°F)
•
Humidity:
0 to 95% RH non-condensing
•
Dimensions (sensor unit):
3.1 x 3.0 x 1.8 inches
(80 x 77 x 46 mm)
•
Flat cable length: 3
7.40
inches (
9
50 mm)
•
IP44 protection, plastic casing (ABS, Polycarbonate) with
PTFE filter
•
Operating environment
•
0°C to 50°C (32° to 122°F)
•
0 to 90% RH non-condensing for indoor use
Contact your Thermo Fisher Scientific representative for
details regarding sensor accuracy, drift and calibration.
Placing the Sensor and
Smart-Vue Pro Data
Logger
Follow these instructions to assemble and place your triple
sensor unit.
1. Place the sensor unit upright inside the chamber to be
monitored being sure to place the flat ribbon cable flush
with the door joint.
Note:
Your cable may be different than that shown in the
following figure.
Figure 24. Ribbon cable lays flat beneath door joint
2. Ensure that the sensor unit stays in an upright position.
Figure 25. Sensor unit inside incubator
CAUTION:
CO
2
detector opening is not a button
and therefore should not be pressed with fingers
or other objects.
Don’t press the
CO
2
detector
opening
Temperature
and humidity
sensors