13
Maintenance
66
Smart2Pure
Thermo Scientific
Replacing the RO/pretreatment
7
Put the cover for the ultrapure cartridge back in place and
switch the system on (connect the main power supply).
9
Discharge the first 5 liters.
Only replace the RO/pretreatment cartridge when the maximum limit for the
feedwater has been exceeded, when you cannot extract any water from the sampling
tap or when the replacement interval is due.
Step
Action
Figure
Step
Action
Figure
1
Switch the system off (unplug main power supply).
2
Remove the cover over the RO/pretreatment cartridge by
pressing the snapper and pull the cover forward.
3
Press on the three quick connectors on the cover of the
RO/pretreatment cartridge and remove the used
cartridge from the system.
4
Locate the new RO/pretreatment cartridge and insert the
cartridge into the system.
Snapper
Cover
Quick
connectors
RO/pretreat-
ment cartridge
Summary of Contents for Barnstead Smart2Pure 12 UF
Page 10: ...Contents viii Smart2Pure Thermo Scientific ...
Page 14: ...1 Transport and packaging 4 Smart2Pure Thermo Scientific ...
Page 18: ...2 Safety Precautions 8 Smart2Pure Thermo Scientific ...
Page 24: ...3 Extend of delivery 14 Smart2Pure Thermo Scientific ...
Page 26: ...4 Intended Use of the device 16 Smart2Pure Thermo Scientific ...
Page 56: ...7 Installation Connections of the Smart2Pure 12 system 46 Smart2Pure Thermo Scientific ...
Page 58: ...8 Flow chart 48 Smart2Pure Thermo Scientific ...
Page 60: ...9 How the Smart2Pure system functions 50 Smart2Pure Thermo Scientific ...
Page 66: ...11 Operating Elements 56 Smart2Pure Thermo Scientific ...
Page 88: ...14 Waste disposal 78 Smart2Pure Thermo Scientific ...
Page 98: ...17 Consumables 88 Smart2Pure Thermo Scientific ...
Page 99: ...Thermo Scientific Smart2Pure 89 18 Terminal assignment ...
Page 100: ...18 Terminal assignment 90 Smart2Pure Thermo Scientific ...
Page 102: ...19 Maintenance record 92 Smart2Pure Thermo Scientific ...
Page 107: ...Thermo Scientific Smart2Pure 97 I ...
Page 108: ...Index ...