3.
Remove each fuse from its fuse holder and inspect it for damage. Carbon
typically coats the inside of failed fuses.
4.
Replace each failed fuse with a 1.6A, Time-Lag T, 250VAC, 5 × 20-mm Fuse.
Note:
The voltage and amperage ratings are on the fuse holder.
5.
Install the fuse holder.
6.
Plug in, then power on the QuantStudio
™
3D Digital PCR Instrument. The
installation is successful if the instrument powers on.
Note:
Fuse failure can result from fluctuations in the supplied power to the
QuantStudio
™
3D Digital PCR Instrument. To prevent further failures, consider
installing an electrical protective device, such as a UPS or surge protector.
Calibrate the touchscreen
Calibrate the instrument touchscreen periodically as part of the routine maintenance
of your instrument, if you observe a decrease in screen responsiveness over time, or
if prompted after updating the instrument firmware.
1.
From the Main Menu of the touchscreen, touch
, then touch
Maintenance &
Service
.
2.
In the Maintenance & Service screen, touch
Screen Calibration
.
Appendix C
Maintenance
Calibrate the touchscreen
C
QuantStudio
™
3D Digital PCR System User Guide
91