Antaris II and Antaris MX Site and Safety Information
49
Specifications
Electrical service
specifications
The following table lists the specifications for electrical service. Contact
your local Thermo Fisher Scientific service representative if you have
questions about the requirements. If you are not sure that your power
lines meet these requirements, contact our customer support for
information about power audits. Locate your power source within 3 m
(10 ft) of the instrument rear panel.
Item
Specification
Electrical service
3-wire
live
neutral
earth ground
Nominal operating AC
input
90 V to 264 V
Nominal operating
current
1.7 A at 115 V
1.0 A at 230 V
Maximum startup
AC input & current
15 A at 115 V
30 A at 230 V
Line frequency
47 Hz to 63 Hz
Line disturbances
Sags, surges or other line disturbances
must not exceed 10 % of input voltage
(even for a half cycle)
Noise
< 2 V (common
mode)
< 20 V (normal mode)
System components such as the computer, printer and monitor are not
included in these specifications. See the documentation that came with
those components for specifics.