Regulatory and Safety Information
Electrical service
Check the line voltage and frequency at the wall outlet or other AC source
that will be used for the instrument before you connect the instrument or
turn on the power. This should be done when the system is installed and
any time the system is moved or connected to another AC source. Specific
service requirements are included in the section entitled “Specifications”,
later in this guide.
Danger
To avoid injury, only a qualified person using the appropriate measuring
device should check the line voltage, current and line frequency.
Power supplies
These FT-NIR instruments receive power from a power supply, which is
mounted on the back panel of the analyzer and can adjust to a variety of
AC voltages (90 – 246 VAC) and line frequencies (47 – 63 Hz)
automatically.
Warning
Avoid shock and fire hazards. Always use an exact replacement for the
fuses, bulbs, and power supplies.
Power cords
You must be sure to use an appropriate power cord for your specific
electrical service. The power cord supplied with the analyzer is a 3-wire,
grounded power cord, appropriate for use in the country listed as the
shipping destination for the system. If the power cord you received is not
appropriate for the electrical system at your site, contact Thermo Fisher
Scientific Customer Support or your local Thermo Fisher Scientific office
and order a new power cord. The illustrations on the next page show the
power cord styles that are available from us.
If the power cord becomes damaged, replace it. Contact technical support
if you need additional information about replacement cord specifications.
Warning
Prevent shock hazard. Do not remove or defeat the ground prong on the
power cord. Always use a grounded outlet.
14
Antaris II and Antaris MX Site and Safety Information