Utility Requirements
Once the line is installed, attach:
•
a shutoff valve
•
a 1/4-inch male fitting or
a 3/8-inch female fitting
The rest of the purge equipment (including the reducing nipple, pressure
coupling, pressure regulator, and flow meter) will arrive with the
instrument. The illustration below shows how the equipment will be
connected between the pressure coupling and the instrument.
If you installed a 1/4-inch male fitting, when your instrument arrives, you
will need a 1/4-inch open-end wrench to attach the pressure coupling. If
you used a 3/8-inch female fitting, you will need a 11/16-inch open-end
wrench to tighten a reducing nipple and a 3/4-inch open-end wrench to
tighten the pressure coupling. Always use pipe tape to prevent leaks at the
fittings.
Shutoff valve
Pressure regulator
Pressure
coupling
3/8" to 1/4"
male reducing nipple
(if required)
Flowmeter
Flow coupler
insert connected
to spectrometer
rear panel
50
40
30
20
10
SCFH AIR
80
100
120
140
160
0
20
40
60
Filter
46
Antaris II and Antaris MX Site and Safety Information
Thermo Fisher Scientific