Thermo Scientific Touch Screen Drybath

Dry Bath with Advanced Microprocessor Control

Catalog Number:
88870007, 88870008
88870009, 88870010
88870011, 88870012
Operation Manual 0000849

Summary of Contents for 88870007

Page 1: ...arranty www thermoscientific com labwarranty User Manual Thermo Scientific Touch Screen Drybath Dry Bath with Advanced Microprocessor Control Catalog Number 88870007 88870008 88870009 88870010 88870011 88870012 Operation Manual 0000849 ...

Page 2: ... Drybath is used in a manner not specified by the manufacturer the protection provided by the unit may be impaired Thank you for selecting Thermo Scientific for your equipment needs Your Thermo Scientific Touch Screen Drybath has been designed with functionality reliability and safety in mind It is your responsibility to use this instrument in conformance with local electrical codes It is very imp...

Page 3: ...eeds for spare or replacement parts or provide you with on site service We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products Whatever Thermo Scientific products you need or use we will be happy to discuss your applications If you are experiencing technical problems working together we will help you locate the problem and chances are correct it yours...

Page 4: ...ion Work 5 1 Operation Guidance 6 1 Examination Before Start Up 6 1 Start Up 6 1 Operation Instructions 6 1 Failure Analysis and Troubleshooting 7 1 Section 1 Section 2 Section 3 Section 4 Section 5 Section 6 Section 7 Section 8 Section 9 Section 10 Modular Block Accessories 8 1 Wiring Diagram 9 1 Warranty Information 10 1 ...

Page 5: ......

Page 6: ...If the plug cannot be plugged into the power socket please ask the electrician to install a correct power socket so as to make the grounded plug to work for safety b Servicing and Replacement of Parts The operator should not take apart the instrument without permission Only qualified professionals are allowed to replace elements or adjust components inside the instrument Replacement of components ...

Page 7: ... they should not be blocked or covered When a single set of instrument is used the interval between ventilation opening before and after the instrument and its nearest object should not be less than 25cm Also don t use the instrument on loose or soft surface or the air inlet of instrument bottom might be blocked Excessive temperature will impair performance or result in failure of instrument This ...

Page 8: ...on the instrument The metal part module near the sign should not be touched with any part of the body when the instrument is operating in a high temperature state or just finished operation to avoid burns Explanation of symbols Attention read user manual before use Alternating Current Protective Conductor Terminal In any of the following cases immediately turn off the power supply withdraw the pow...

Page 9: ......

Page 10: ...Fisher Scientific will not be held responsible for damages The product is dry bath with advanced microprocessor control which can be widely applied to sample reservation enzyme reservation and reaction DNA amplification electrophoresis degeneration and serum coagulation etc The characteristics of the product are as follows The digital display and control of the temperature Heat blocks help avoid p...

Page 11: ......

Page 12: ...nsportation and storage conditions Ambient temperature 20 C 55 C Relative humidity 80 3 Basic parameters Table 1 100 120V Parameters Type Touch Screen 1 block 100 120V Touch Screen 2 block 100 120V Touch Screen 4 block 100 120V Temperature Range Ambient Temperature 5 C 130 C ambient temperature 25 C Temperature Uniformity 1 0 C Temperature Accuracy 0 5 C 37 C Temperature Fluctuation 0 5 C Temperat...

Page 13: ... 4 Features Table 2 220V Parameters Type Touch Screen 1 block 200 240V Touch Screen 2 block 200 240V Touch Screen 4 block 200 240V Temperature Range Ambient Temperature 5 C 130 C ambient temperature 25 C Temperature Uniformity 1 0 C Temperature Accuracy 0 5 C 37 C Temperature Fluctuation 0 5 C Temperature Rise Time 20 min rise from 30 C to 130 C Size L W H 373 200 100 mm 14 7 7 87 3 94 in 403 200 ...

Page 14: ...time 1 Structure Interface Figure 5 1 Unit Front Figure 5 2 Unit Back Housing Front Cover LCD Touch Screen Outlet for External Temperature Sensor Product Label Handle Power Socket Fuse Housing Power Switch Note 1 When the external temperature sensor is connected the internal temperature sensor will disconnect automatically and the temperature control and display will relay on the external temperat...

Page 15: ......

Page 16: ...rmal after startup turn off the power source immediately and contact Thermo Fisher Scientific Press the power switch and the instrument will be powered on The LCD screen of the operation panel will show the main interface The main interface Press Single point set to enter into the single setting interface Single setting interface Press BACK to go back to the main interface Press RUN to enter into ...

Page 17: ...erature stabilizes for 30 minutes b Calibrate the instrument with the qualified Grade II standard filled thermometer c Calibration point The middle hole of the heat block Pour paraffin oil into the well and soak the thermometer bulb Instruction During the operation of the instrument the temperature and the time should not be changed If it they should stop the program and enter into the setting int...

Page 18: ...ss YES and the program will run in the changed form but the changes won t be saved Press NO to go back to the program setting interface The user can set 1 5 sections of the program Press to delete the corresponding section when there is only one section this step would be of no use Press to add sections Choose the corresponding box to set the corresponding temperature and time Caution This instrum...

Page 19: ...CAL and the temperature is unable to be adjusted Refer to Single Running Interface Press SKIP to skip to the next section If it is the last section the interface of asking you whether to stop the program will appear Refer to Single Running Interface Press PAUSE the program will be paused and the temperature will be maintained to the set temperature and the pause interface will pop up Press STOP th...

Page 20: ...use Damage of the switch Replace the switch Others Contact Thermo Fisher Scientific 2 The LCD shows Err code 0x1 and the buzzer alarms in a di di sound Short circuit of sensor Check whether the connecting wire of sensor is damaged and replace it 3 The LCD shows Err code 0x2 and the buzzer alarms in a di di sound Open circuit of sensor Check whether the connecting wire of sensor is damaged and repl...

Page 21: ...st tubes 32 holes for 6mm test tubes 21 holes for 10mm test tubes 32 21 88870112 For mixed size test tubes 18 holes for 1 5ml test tubes 10 holes for 2 0ml test tubes 18 10 88870113 For mixed size test tubes 3 holes for 25mm test tubes 12 holes for 13mm test tubes 6 holes for 6mm test tubes 3 12 6 88870114 For mixed size test tubes 30 holes for 0 5ml test tubes 20 holes for 0 2ml test tubes 30 20 ...

Page 22: ...Drybath Incubator 9 1 Thermo Scientific Section 9 Wiring Diagram The page is just for reference and is subject to change without prior notice ...

Page 23: ......

Page 24: ...o s option all non conforming parts must be returned to Thermo Fisher Scientific postage paid and replacement parts are shipped FOB desti nation THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER WRITTEN ORAL OR IMPLIED NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY Thermo shall not be liable for any indirect or consequential damages including ...

Page 25: ......

Page 26: ...anada 1 866 984 3766 866 9 Thermo www thermo com Europe Austria 43 1 801 40 0 Belgium 32 2 482 30 30 France 33 2 2803 2180 Germany national toll free 08001 536 376 Germany international 49 6184 90 6940 Italy 39 02 02 95059 434 254 375 Netherlands 31 76 571 4440 Nordic Baltic countries 358 9 329 100 Russia CIS 7 812 703 42 15 Spain Portugal 34 93 223 09 18 Switzerland 41 44 454 12 12 UK Ireland 44 ...