4
Rotor Loading
Improper Loading
Thermo Scientific
75006476 LH-4000W
4-3
Improper Loading
Maximum Loading
The rotor can run at high speeds. The rotor design has sufficient reserve stability even when spinning at
top speed.
The safety system of the centrifuge requires that you do not overload the rotor.
There are two options available for centrifuging samples whose weight, including adapt
e
r, exceeds the
maximum permissible load:
•
Reduce the fill level.
•
Reduce the speed.
Use the table or the formula:
Actual Load
maximum speed
600
6000
620
5902
640
5809
660
5721
680
5636
700
5555
720
5477
740
5403
760
5331
780
5262
800
5196
820
5132
840
5071
860
5012