5
Aerosol-tight Applications
Fill Level
Thermo Scientific
75006476 LH-4000W
5-3
To achieve a uniform pressure onto the seal
,
turn the cap another 1
½
grip areas (some 15°).
Use the markings on the bucket for orientation.
Fill Level
The tubes are only to be filled to a level which ensures that the sample is unable to reach the top of the
tube during centrifugation. Therefore
,
fill the tube only 2/3 of the rated level.
Checking the Aerosol-Tightness
The aerosol tightness testing of the buckets depend
s
on the microbiological test process in
accordance with the EN 61010-2-020 Appendix AA.
Whether or not a bucket is aerosol-tight depends primarily on proper handling.
Check as needed to make sure your bucket is aerosol
-
tight.
The careful inspection of the seals and seal surfaces for signs of wear and damage such as cracks,
scratches and embrittlement is extremely important.
Aerosol-tight applications are not possible if the lids are open.
Aerosol-tightness requires the correct operation when filling the sample vessels and closing the bucket
cap.
Grip Area
Marking
Round Bucket 75006477
After greasing the seal turn the cap until it fits slightly on the bucket.