6
Maintenance and Care
Decontamination
6-4
75006476 LH-4000W
Thermo Scientific
4.
Remove the buckets from the rotor
.
5.
Remove the centrifuge tubes and adapters from the bucket and disinfect them
.
6.
Release the rotor
.
7.
Grasp the rotor with both hands and lift it vertically off the centrifuge spindle
.
8.
Treat the rotor and accessories according to the instructions for the disinfectant (spray or soak in
solution). Adhere strictly to the given application times
.
9.
Be sure the disinfectant can drain off the buckets
.
10.
Rinse the rotor and accessories thoroughly with water
.
11.
Dispose of the disinfectant according to the applicable guidelines
.
12.
Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50
°C
.
13.
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil
(70009824). Also treat the cavities with oil
.
14.
Trea
t
the bol
t
s with bol
t
grease (75003786).
Decontamination
Decontaminate rotor and centrifuge immediately whenever radioactive material has spilled during
centrifugation.
For general radioactive decontamination use a solution of equal parts of 70
% ethanol, 10
% SDS
and water.
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Remove the buckets from the rotor.
5. Remove the centrifuge tubes and adapters from the bucket and decontaminate them.
6. Release the rotor.
7. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
WARNING
Radioactive material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if need be.
CAUTION
Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.