Rotor Installation
Rotor Installation
3-2
75006476 LH-4000W
Thermo Scientific
Rotor Installation
Use only rotors with approved accessories.
Proceed as follows:
1.
Open the lid of the centrifuge and if necessary
,
remove any dust, foreign objects or residue from the
chamber
.
Thread and
O
-ring must be clean and undamaged
.
2.
Hold the rotor over the centrifuge spindle
.
3.
Let the rotor slide slowly down the centrifuge spindle
.
4.
Thread the fastening tool into the centrifuge spindle clockwise. Hold the rotor with the other hand
into position
.
5.
Check if the rotor is properly installed by lifting it slightly on the handle
.
6. If available close the rotor with the rotor lid.
Removing the Rotor
To remove the rotor, proceed as follows:
1.
Open the centrifuge door
.
2.
Thread the fastening tool in the centrifuge spindle counterclockwise. Hold the rotor with the
other hand into position
.
3.
At the same time, pull the rotor directly upwards with both hands and remove it from the
centrifuge spindle. Make sure not to tilt the rotor while doing this
.
CAUTION
Unapproved or incorrectly combined accessories can cause serious damage to
the centrifuge.
CAUTION
Do not force the rotor onto the centrifuge spindle.
WARNING
Check for any damage to the rotor: Damaged rotors must not be used.
Keep the centrifuge spindle area of the rotor clear of objects.
CAUTION
Check that the rotor is properly locked on the centrifuge spindle before each
use by pulling it at its handle.
Be sure to check all sealings before starting any aerosol-tight applications.