1
Introduction and Description
Technical Data
Thermo Scientific
Heraeus
Multifuge X3F / X3FR
1-3
Technical Data
The technical data of the Heraeus Multifuge X3 F / X3 FR is listed in the following table.
Table 1-1.
Technical Data Heraeus Multifuge X3 F / X3 FR
Feature
Value
Environmental conditions
-Use in interior spaces
-Altitudes of up to 2,000 m above sea level
-max. relative humidity 80% up to 31 °C; decreasing linearly up to 50%
relative humidity at 40 °C.
permissible ambient temperature
+2 °Cto +35 °C
+2 °C to +35 °C
Overvoltage category
II
II
Pollution degree
2
2
Heat dissipation
ventilated
refrigerated
230
V
120
V
230
V
120
V
6653 BTU/h
4776 BTU/h
5800 BTU/h
4776 BTU/h
IP
20
20
running time
unlimited
unlimited
max Speed n
max
15200 rpm (depending on the rotor)
15200 rpm (depending on the rotor)
min Speed n
min
300 rpm
300 rpm
maximum RCF value at n
max
25314 x g
25314 x g
maximum kinetic energy
< 62.5 kJ
< 62.5 kJ
noise level at maximum speed
< 64 dB (A)
< 64 dB (A)
temperature setting range
-10 °C to +40 °C
Dimensions
ventilated
refrigerated
Height
Width
Depth
Table top height
840 mm
550 mm
670 mm
790 mm
840 mm
550 mm
670 mm
790 mm
weight without rotor
152.5 kg
142.5 kg
Thermo Scientific
Heraeus Multifuge
X3 FR
Refrigerant
Refrigerant
Quantity
Pressure
GWP CO2e
75004559
R-404A
0.54 kg
31 bar
3922
2.12 t
Contains fluorinated greenhouse gases in a hermetically sealed system.