29 cu ft Environmental Chamber 3-3
Thermo Scientific
Section 3
Configuration
On units that have the RS485 option, direct communication with the
Model 1535 alarm system can be established. Each piece of equipment
connected to the 1535 must have a unique address. An address of 0-24
can be entered for the incubator. A setting of 0 is an invalid address that
the 1535 will ignore. The factory setting for the RS485 address is 0.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until RS485 XX is displayed in the message
center.
3. Press up/down arrow to move the RS485 address.
4. Press Enter to save the RS485 address.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
Communications
Address - RS485
Summary of Contents for 3960
Page 14: ...29 cu ft Environmental Chamber 1 7 Thermo Scientific Section 1 Installation and Start Up ...
Page 15: ......
Page 18: ...29 cu ft Environmental Chamber 2 3 Thermo Scientific Section 2 Calibration ...
Page 19: ......
Page 23: ...3 4 29 cu ft Environmental Chamber Thermo Scientific Section 3 Configuration ...
Page 29: ......
Page 37: ......
Page 39: ...8 2 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Page 42: ...29 cu ft Environmental Chamber 8 5 Thermo Scientific Section 8 Parts ...
Page 43: ...8 6 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Page 44: ...29 cu ft Environmental Chamber 8 7 Thermo Scientific Section 8 Parts ...
Page 45: ......
Page 46: ...29 cu ft Environmental Chamber 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Page 47: ...9 2 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Page 48: ...29 cu ft Environmental Chamber 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Page 49: ...9 4 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Page 50: ...29 cu ft Environmental Chamber 10 1 Thermo Scientific Section 10 Appendix ...
Page 51: ......