29 cu ft Environmental Chamber 3-1
Thermo Scientific
Section 3
Configuration
Several features available in the Configuration Mode allow custom setup of
the incubator. These features are listed and described below. All features
may not be necessary in all applications, but are available if needed. To
enter Configuration mode, press the Mode key until the Config indicator
lights. Press the right and/or left arrow until the appropriate parameter
appears in the message center. See Chart 3-1 at the end of this section for
more detail.
The audible alarms can be turned on or off. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Audible XXX is displayed in the message
center.
3. Press up/down arrow to toggle Audible ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to run mode or right/left to go to
next/previous parameter.
A 3-digit Access Code can be entered to avoid unauthorized personnel
from changing the setpoints, calibration, or configuration. A setting of
000 will bypass the access code. The factory setting is 000.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Acc Code XXX is displayed in the message
center.
3. Press up or down arrow to change the access code.
4. Press Enter to save the access code.
5. Press the Mode key to return to the Run mode or right/left to go to
next/previous parameter.
Turn All Audible
Alarms On/Off
Set an Access Code
Summary of Contents for 3960
Page 14: ...29 cu ft Environmental Chamber 1 7 Thermo Scientific Section 1 Installation and Start Up ...
Page 15: ......
Page 18: ...29 cu ft Environmental Chamber 2 3 Thermo Scientific Section 2 Calibration ...
Page 19: ......
Page 23: ...3 4 29 cu ft Environmental Chamber Thermo Scientific Section 3 Configuration ...
Page 29: ......
Page 37: ......
Page 39: ...8 2 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Page 42: ...29 cu ft Environmental Chamber 8 5 Thermo Scientific Section 8 Parts ...
Page 43: ...8 6 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Page 44: ...29 cu ft Environmental Chamber 8 7 Thermo Scientific Section 8 Parts ...
Page 45: ......
Page 46: ...29 cu ft Environmental Chamber 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Page 47: ...9 2 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Page 48: ...29 cu ft Environmental Chamber 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Page 49: ...9 4 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Page 50: ...29 cu ft Environmental Chamber 10 1 Thermo Scientific Section 10 Appendix ...
Page 51: ......