3-2
29 cu ft Environmental Chamber
Thermo Scientific
Section 3
Configuration
Set Low Temp Alarm
Limit (Tracking Alarm)
The low temp alarm limit is the deviation from the temperature setpoint,
which will cause a low temp alarm. The low temp alarm is variable from
0.5° below setpoint to 5.0° below setpoint. The factory setting is 1.0°
below setpoint. A minus sign in the display indicates that the alarm setting
is below the setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Temp Lo -X.X is displayed in the message
center.
3. Press up/down arrow to change the low temp alarm limit.
4. Press Enter to save the low temp alarm limit.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
The temperature alarms can be programmed to trip the remote alarm
contacts. A setting of ON will cause this, a setting of OFF will not allow
temperature alarms to trip the contacts. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until TMP RLY XXX is displayed.
3. Press the up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run or the right/left arrow key to go
to next/previous parameter.
Enable Temp Alarms
to Trip Relay Contacts
Summary of Contents for 3960
Page 14: ...29 cu ft Environmental Chamber 1 7 Thermo Scientific Section 1 Installation and Start Up ...
Page 15: ......
Page 18: ...29 cu ft Environmental Chamber 2 3 Thermo Scientific Section 2 Calibration ...
Page 19: ......
Page 23: ...3 4 29 cu ft Environmental Chamber Thermo Scientific Section 3 Configuration ...
Page 29: ......
Page 37: ......
Page 39: ...8 2 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Page 42: ...29 cu ft Environmental Chamber 8 5 Thermo Scientific Section 8 Parts ...
Page 43: ...8 6 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Page 44: ...29 cu ft Environmental Chamber 8 7 Thermo Scientific Section 8 Parts ...
Page 45: ......
Page 46: ...29 cu ft Environmental Chamber 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Page 47: ...9 2 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Page 48: ...29 cu ft Environmental Chamber 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Page 49: ...9 4 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Page 50: ...29 cu ft Environmental Chamber 10 1 Thermo Scientific Section 10 Appendix ...
Page 51: ......