1500 Series A2 4-11
Thermo Scientific
Section 4
Operation
Figure 4-15 shows that it is in Idle mode with the window ready. In this
case, the downflow and inflow status indicator boxes show an “Alert” icon.
Figure 4-15.
Main Screen - Unit in Idle Mode, window ready
Figure 4-16 shows that it is in Idle mode with the window open. In this
case, the window status indicator boxes show an “Alert” icon.
Figure 4-16.
Main Screen - Unit in Idle Mode, window opened
The biological safety cabinet has a built-in UV disinfection routine that
affords an intensifying additional disinfection after a standard wipe/spray
disinfection. UV disinfection can be started by pressing the “UV” button
on the Control Panel or can be started automatically with the specified
time if the UV-auto-start function is enabled.
For running the UV disinfection routine, the front window must be
completely lowered to protect against UV radiation. A safety interlock
ensures that the routine cannot be run until the front window is in the
closed position. And the LED light must be switched off.
Idle Mode (continued)
Running UV
Disinfection