1500 Series A2 1-5
Thermo Scientific
Section 1
Description
The sample chamber is accessible through various positions of the front
window. The working position of the window is a 10 inch opening,
+0.25”, -0.75” (depending on model - check data label and first page of
this manual). For loading the chamber, the window opens fully. For
decontamination and energy saving, enter into Night-Set-Back mode by
closing the window. When the window is closed, the blower speed is
reduced, minimizing energy consumption and noise emittance without
compromising particulate containment inside the work chamber.
Note
The working and loading window positions are indicated on the user
interface, as controlled by micro switches and displayed by status
indicators on the control panel.
s
Use of the Window
Window
Window
Pillardress
panel
Opening height for
cleaning window
Pillardress
panel
Closed position
Work
position
Max opening for
loading samples
Figure 1-3.
Window Openings