Nanopure™ Analytical 10-1
Thermo Scientific
Section 10
Accessories
Accessories D8964 (float switch) and D2706 (pressure switch) are designed
to protect the Nanopure Analytical pump by alerting the system of an
inadequate feed water condition so that the pump can be shut down. If an
inadequate feed water condition exists and the pump is shut down due to
this condition, the display will read “Check Inlet.” Follow the proceeding
instructions for installation.
1. Disconnect the unit from the electrical power.
2. If using the D8964 float switch, follow the installation instructions
included with the float switch for installation into a tank.
3. If using the D2706 low pressure switch, install the PVC tee (supplied
with D2706) in the incoming water line. Screw the switch into the top
of the tee, then connect the inlet tubing of the Nanopure Analytical
into the remaining opening of the PVC tee.
4. Route the cable from the float or low pressure switch to the top left of
the Nanopure Analytical.
Feedwater Float or
Pressure Switch
Tank
D8964
Float Switch
D2706
Pressure
Switch
Figure 8-1.
Feedwater Float/ Pressure Switch