Nanopure™ Analytical 7-3
Thermo Scientific
Section 7
Initial Operation
Initial Rinse
(continued)
10. Unit will display, “Air Purge?” Press ENTER.
11. Upon completion of the air purge, complete the “System Cleaning
Procedure.”
Caution
Never dispense water without first installing hose barb provided in
parts bag or final filter supplied with cartridge pack.
s
Your Nanopure Analytical has been shipped with a cleaning syringe (Cat.
No. CMX25). When the system is new or whenever a new cartridge pack
is installed, the system should be cleaned. It is best to perform cleaning
procedures when the system or water is not required. The entire cleaning
procedure will take approximately 3.5 hours to complete. Clean your
Nanopure as follows:
1. (If you have installed an Accudispense volumetric remote dispenser,
place it in Manual Mode.) From the (Idle) display, press the UP or
DOWN arrow until display reads, “Clean Unit?”
2. Press ENTER.
3. Unit display will ask if you have “Disconnected Inlet Water (NO).”
4. Disconnect feed water supply at the quick-disconnect inlet valve on
left side of unit.
5. Press the UP or DOWN ARROW to select (YES) and then press
ENTER. Display will now read,
“Injected Cleaner (NO).”
6. Open front (left side door) to
expose cartridge pack. Remove
solution syringe from wrapper.
7. Remove luer cap on system
injection port, located on the
right side of the cartridge pack
manifold by turning
counterclockwise.
8. Remove syringe luer cap and
attach syringe to the system luer
fitting.
System Cleaning
Procedure
Injection Port
Figure 7-2.
Injecting Cleaning Solution