Nanopure™ Analytical 5-1
Thermo Scientific
Section 5
Installation
The Nanopure Analytical features a removable control panel display which
allows the system to be mounted almost anywhere within the laboratory.
Use the wall bracket for wall mounted systems as a template to drill
mounting holes. (The Nanopure Analytical does not include screws and
fasteners for mounting.) Please refer to the “General Specifications” section
for clearance requirements.
Note
The outlet of a gravity feed storage reservoir must be above or at the
same level as the inlet of the Nanopure Analytical.
s
Caution
Wall composition, condition and construction as well as fastener
type must be considered when mounting this unit. The mounting surface
and fasteners selected must be capable of supporting a minimum of 150
lbs.(68kg). Inadequate support and/or fasteners may result in damage to
mounting surface and/or equipment. If you are unsure of mounting
surface composition, condition and construction or correct fasteners,
consult your building maintenance group or contractor.
s
Warning
Do not locate the Nanopure Analytical directly over equipment
that requires electrical service. Routine maintenance of this unit may
involve water spillage and subsequent electrical shock hazard if improperly
located.
Warning
Do not use in the presence of flammable materials; fire or
explosion may result. This device contains components which may ignite
such materials.
s