Method:
1. Lower the front window to a height of 75 mm.
2. Operate the system blowers for approximately 20 minutes.
3. Insert the thermal anemometer probe into the probe holder. Place the
probe holder at the window opening work surface, ensuring it is seated
properly. See Figure 8-1. Adjust the probe so the air passing through
probe is centered in the window opening (37.5 mm).
4. Beginning 150 mm in from the
left inside wall, record
successive readings every
150 mm (7 readings for 1.2 m
models and 11 readings for
1.8 m models).
5. Average those readings and
calculate the inflow velocity
(V1) as described below.
Calculation:
• Inflow velocity (V1) = average of
actual inflow velocity * factor *
0.3
Acceptance:
• 0.5 m/s, ±0.025 m/s
Note
The thermoanemometer
probe must be positioned at a 10°
angle from front vertical and
located on the same plane as the
inside of the window glass.
Multiply the K factor (0.9929 for
1.2 m, 0.9918 for 1.8 m) by the
average velocity reading (V1).
Refer to the data plate on the lower
right corner of the unit face.
s
1300 Series A2
8-7
Thermo Scientific
Section 8
Certification Testing
Testing Information
(continued)
Figure 8-1.
Constricted Window Method
Summary of Contents for 1300 Series
Page 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Page 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Page 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Page 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Page 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Page 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Page 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...