Checking these items on a routine basis, will help to ensure your
cabinet is always safe and operating as intended.
Warning
The diffuser plate on the chamber ceiling serves as protection for
the downflow filter and prevents refluxing of downflow air. When
scanning the filter surface for a leak test, the perforated plate must be
removed.
s
As filter replacement is an interference with the safety system of the unit,
filters must only be replaced by adequately trained and authorized service
personnel.
Warning
For any service work in a potentially contaminated section of the
biological safety cabinet, the unit must first be decontaminated.
s
The sample chamber is illuminated by two fluorescent bulbs. The spring-
loaded sockets of the fluorescent bulbs are installed in the light mounting
frame in the sample chamber (Figure 6-1).
1. Switch the unit off and disconnect it from the power source.
2. Move the window to the cleaning position (below closed position) to
ensure a sufficient gap between the window’s upper edge and the light
dome.
3. The fluorescent bulbs are installed in rotating sockets. To remove,
rotate the bulb counterclockwise to disengage the latch and remove it
from the sockets.
4. To install, slide the bulb contact pins into the rotating socket grooves
and rotate the bulb clockwise to latch the sockets.
The optional, factory installed UV light is installed in the chamber ceiling
immediately behind the window. The UV light should be replaced after
1500 operating hours. See Figure 6-1.
1. Turn the unit off and disconnect it from the power source.
2. Move the window to the maximum open position.
6-2
1300 Series A2
Thermo Scientific
Section 6
Maintenance
Service
Sample Chamber Lights
Optional UV Lights
Summary of Contents for 1300 Series
Page 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Page 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Page 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Page 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Page 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Page 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Page 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...