1. Rinse all surfaces
twice
with clean distilled water.
2. Discard any liquid from the drain pan. Rinse pan.
3. Dry all chamber surfaces completely.
UV disinfection can be performed by using the optional factory installed
UV light.
To start the UV disinfection procedure;
1. Close the front window, the air system operates in the reduced mode
(blue LED illuminated).
2. Press and hold the UV key on the control panel until the display
alternates between dIS and the remaining disinfection time.
To interrupt or cancel the UV disinfection procedure, just press the UV
key (the display shows the operating hours) and slide the window up.
The UV disinfection time is preset. However, this setting can be changed,
if required.
1. Turn the unit on and move the window to the work position.
2. Press and hold the UV key until the preset time is displayed.
3. To increase the time, press the internal power supply key. Each time
the key is pressed, the time increases by 30 minutes.
4. To reduce the time, press the sample chamber light key. Each time the
key is pressed, reduces the time by 30 minutes.
5. To save the value, press the UV key. The display shows the operating
hours again.
Note
The UV disinfection time is adjustable from 0:00 (disabled) to 24:00
hours.
s
1300 Series A2 5-3
Thermo Scientific
Section 5
Cleaning and Decontamination
Rinsing
UV Disinfection After
Cleaning
Change the UV Disinfection Time
Summary of Contents for 1300 Series
Page 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Page 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Page 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Page 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Page 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Page 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Page 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...