The canister ring has an adjustable slide feature that will accommodate canisters of 4.75 to
6.25 inches (120-160mm) in diameter. Loosen thumb screw on bottom of ring and slide to
adjust. Tighten knob again when in desired position.
Insert secondary display (optional) into rear of main unit, until end of tube is flush with lower
mount. Tighten (4) mounting screws with alignment tool.
Perform electrical safety tests as required per institutional procedure. These include but are not
limited to: leakage current, high-pot, and ground bond test.
WARNING
Grounding reliability can only be achieved when the power cord is connected to a properly
grounded receptacle. Risk of electric shock exists if the equipment is not connected to a properly
grounded receptacle and can result in death or serious injury to the patient or user.
The Electrical Safety Check must be performed by qualified personnel authorized by the institution
to perform such testing. The Electrical Safety Check must be performed and documented at least
once per year, or according to institutional policy. The degree of protection against electric shock:
Type BF Applied Part.