Section 10 – Troubleshooting Guide
This device contains no user serviceable parts. All service must be performed by an authorized
representative of Thermedx
®
.
D
ESCRIPTION
:
U
SER
S
UGGESTIONS
/
S
OLUTIONS
:
W
HERE
“
CALL FOR SUPPORT
”
IS INDICATED
,
PLEASE CALL YOUR DISTRIBUTOR OR
T
HERMEDX
®
AT
888-542-9276.
F
OR CUSTOMERS OUTSIDE THE
US,
PLEASE
CALL YOUR DISTRIBUTOR OR
T
HERMEDX
®
AT
011.1.440.542.0883.
Device does not
turn on
Check that the Device is fully plugged into the wall, and the power cord is fully
plugged into the unit.
Check that Device’s power switch is on.
Check wall outlet power.
If problem persists call for support.
Device alarms on
start up
Follow on screen instructions for any error messages displayed. See the errors
and alarms section of this manual for further information.
Ensure that the disposable (if one is fitted) is not pressurized. Disconnect from
any surgical instruments and unclamp outflow to relieve pressure.
Ensure the disposable (if one is fitted) is not over-heated from a previous use.
Allow time to cool.
Reboot device.
If problem persists call for support.
No or low flow,
poor visualization
Confirm the
RUN
button has been pressed to begin fluid flow.
Check for empty fluid bags, confirm clamps are open, and look for kinks and
twists in bags and tubing.
Check to make sure inflow and outflow valves are open.
Check for occluded scope.
If flow is present but low, increase pressure setting.
If problem persists call for support.
Hysteroscopic
tissue removal
system
troubleshooting
To increase distention, increase pressure or reduce suction by adjusting regulator.
If no regulator is available, open an extra port on a suction canister to reduce
suction.
To increase suction: turn up vacuum regulator or reduce the number of suction
lines.
If suction is inadequate, go back to system setup screens and view suction
canister setup instructions.
Device has
burning smell
Remove Cartridge, check for debris. Do not reuse cartridge.
Confirm Fluid Bag(s) are not empty. Replace as necessary.
Confirm the vents on the bottom and rear of the unit are not blocked.
If problem persists call for support.
Device is making
noise
A faint whistling noise is normal while priming due to air venting out of the
cartridge.
Repetitive “sucking” noise from the pump is due to lack of fluid. Check fluid levels
and clamps.
Faint “ticking” noises from the fans or pump are indicative of normal operation.
If problem persists call for support.
Device is not
warming fluid
Confirm Warming is enabled. Go to the settings screen to enable warming.
Allow fluid to flow. Warming is enabled only when fluid is flowing.
If the no fluid icon is displayed, ensure tubing is fully seated in the bubble
detector and bubble free.
Ensure no error messages are displayed in the message center on the bottom of
the screen.
If problem persists call for support.