Press the
Close
button to save new canister
weights and leave the screen.
Once back to the Run screen:
reconnect/unclamp vacuum and/or suction
source to resume fluid collection.
Note: If pumping continued during the
change, fluid will have accumulated in drape
or cavity. Initially this will appear as Deficit
and may cause an alarm. Continue
suctioning and deficit volume will decrease as
fluid returns to canisters.
M
ISCELLANEOUS
Low Bags
When a bag’s volume drops below 30%, the
bag indicator background turns red. When
volume drops below 20%, a tone sounds.
User should prepare more bags to continue
past remaining volume. If bags run dry a
“Fluid not Flowing” message will display, and
the unit will continue to pump in an attempt
to maintain pressure.
At the end of a procedure, if the pressure is
too high, relieve the fluid pressure within
cartridge/tubing set by unclamping distal
tubing and draining. Notification will
extinguish once pressure is reduced.