Section 2: Quick Start Guide
15
Establish Reference Cables
Continued. . .
3B. Loopback Reference –
Perform these steps on both Local and
Remote units:
a. From AutoTest Screen press the REF
softkey.
b. Use the UP/DOWN arrow keys to high
light LOOP BACK .
c. Press ENTER button to select.
d. Per the screen diagram, connect the
Reference Cable from the Power Meter
Port to the Source Port.
e. Press the NEXT softkey to continue.
f. When complete, a checkmark will
appear on screen.
4.
Press DONE softkey to return to the AutoTest Screen.
5.
Disconnect the reference cables from each other.
6.
Leave the fiber connected to the Source Port.
Note: Disconnection of this cable will cause loss of references.
You are now ready to perform an AutoTest.