CL-B101D40-TCRTD Manual (d0099)
STEP A Enter the Calibration Mode
1)
Press the
P
and buttons at the same time.
Display toggles between [CAL] and [oFF].
2) Press the or button. Display changes from [oFF]
to [on].
3) Press the
P
button.
STEP B
If the Display toggles between [CAL] and [out] the
optional Analog Output hardware is installed. In which
case select [CAL] [out]. Display toggles between [thEr]
or [rtd] depending on whether a IT10 Thermocouple
Input Module or a IT11 RTD Input Module are detect-
ed by the soft ware. If no optional output hardware is
installed the menu will skip directly to STEP C.
STEP C Sensor Type Selection
1) Using the and buttons, adjust the display
to the desired sensor type.
2) Press the
P
button. Display toggles between
[rES] resolution select and previous [rES] setting.
STEP D Set the Resolution
1) Using the and buttons, adjust the display
to the desired resolution [rES] value.
2) Press the
P
button. Display toggles between
[dEG] and previous [dEG] selection.
STEP E Selection of Degree C or Degree F
1) Using the and buttons, adjust the display to
either ˚C or ˚F.
2) Press the
P
button.
Calibration Procedure
MAIN MENU
Operational Display
Sub-menu
MODE
STEP A Calibration
Mode
STEP B Calibration
Mode
To Digital Span Selection for
Bargraph Display Page 5
cAL out/iP branch will only appear if
the analog output option is installed,
otherwise cAL on goes directly to step c.
To Digital Span Selection
for Bargraph Display
See Page 6
To Two Point Analog
Output Range Setting
and Calibration
See Page 7
[K]
[r]
[t]
If RTD module is installed
this menu appears
3 or 4 wire
RTD input
selection
RTD select type
P385 or P392
If Thermocouple module is installed
this menu appears
Resolution select
Thermocouple select
Select J, K, R or T
STEP C
STEP D
STEP E
1
°
0.1
°
Degree select
°
C
°
F
THERMOCOUPLE OR RTD TYPE,
RESOLUTION AND DEGREE C OR F SELECT
Input Sensor
Reso-
lution
°
C Range
°
F Range
J T/C
K T/C
R T/C
T T/C
100
Ω
RTD (385 curve)
100
Ω
RTD (392 curve)
-120 to 760
°
C
-120.0 to 530.0
°
C
-120 to 1370
°
C
-120.0 to 530.0
°
C
0 to 1760
°
C
0 to 530.0
°
C
-120 to 400
°
C
-120.0 to 400.0
°
C
-200 to 800
°
C
-199.9 to 530.0
°
C
-200 to 800
°
C
-199.9 to 530.0
°
C
-200 to 1400
°
F
-199.9 to 999.9
°
F
-200 to 2500
°
F
-199.9 to 999.9
°
F
32 to 3210
°
F
32.0 to 999.9
°
F
-200 to 750
°
F
-199.9 to 750.0
°
F
-200 to 1470
°
F
-199.9 to 999.9
°
F
-200 to 1470
°
F
-199.9 to 999.9
°
F
Sensor Range Table
1
°
0.1
°
1
°
0.1
°
1
°
0.1
°
1
°
0.1
°
1
°
0.1
°
1
°
0.1
°
Thermocouple type or RTD type selection mode
Input Module Calibration Procedure
See page 12 for the Calibration instructions of each Input Module type.