CL-B101D40-TCRTD Manual (d0099)

101.60

4.000

42.85
1.687

42.85

1.687

42.85

1.687

42.85
1.687

6.50

0.256

113.79

4.480

113.79

4.480

56.90

2.240

56.90

2.240

17.50
0.689

97.77

3.849

119.70

4.713

13.70
0.539

1/4"-28UNF

Panel Cutout

5.0

0.197

R4

0.157

16.30
0.642

13.70
0.539

4.80

0.189

screw-10-32 UNF

8

0.315

7

0.276

16.30
0.642

13.70
0.539

2.60

0.102

6.32

0.249

screw-1/4”-28 UNF

Case Dimensions

IT11

: RTD, 100

Ω

 Pt. Select 

3/4-wire,

°

C/

°

F, 1

°

/0.1

°

IT15

: RTD, 1000

Ω

 Pt. Select 3/4-wire, 

°

C/

°

F, 1

°

/0.1

°

Lead Resistance Compensation

RTD

Pt-100

Ω

RTD

3 wire

4 wire

Excitation is 1mA. Up to 50

Ω

 resistance 

in each lead can be compensated

Typical accuracy is

±

(0.05% + 2 digits)

LINEARISATION IS DIGITAL

LEOPARD

167G

BL-40H
DL-40H
FL-B101D40H

IT10 Thermocouple Input Signal Conditioner installed.

1�  The cold junction select header must be installed in the 

correct position for the thermocouple type to be used� 

Thermocouple types J, K, R and T are supported� If you 

wish to use a different thermocouple from the default setting 

of K/T it is necessary to remove the module and move the 

cold junction select header to the appropriate position�

2�  Unplug the connector plugs from the meter� Remove the 

case back panel  and slide the module out of the case�

3.   After  selecting  the  appropriate  header  position,  insert  the  

module back into the case� Snap the back panel back into 

the case� Apply power to the meter�

4�  Enter the program mode and select the type of thermocouple 

(J, K, R, T), the resolution (0�1° or 1°) and the display units 

·°C or °F)� See Page 4 of the data sheet for details�

5� Connect a thermocouple simulator to the meter inputs� 

Apply an input corresponding to 0° and adjust the ZERO 

Potentiometer to make the display read 0�

6.  Apply an input corresponding to the maximum reading of 

the thermocouple and adjust the SPAN Potentiometer to 

make the display read correctly�

7�  The meter is now calibrated and ready for use� Calibration 

will have to be performed again if the thermocouple type is 

changed�

IT11 or IT15 RTD Input Signal Conditioner installed.

1.  Enter the program mode and select the type of RTD (385 

or  392  curve    and  3-wire/  4-wire),  the  resolution  (0.1°  or 

1°) and the display units ·°C or °F)� See Page 4 of the data 

sheet for details�

2�  Connect an RTD simulator to the meter inputs� Apply an input 

corresponding to 0° and adjust the ZERO Potentiometer to 

make the display read 0�

3.  Introduce a lead resistance of 10Ω in each lead. Adjust the 

Lead Resistance Compensation potentiometer to make the 

display again read 0�

4�  The meter is now calibrated and ready for use� Calibration 

will have to be performed again if the RTD type is changed�

Input Module Calibration Procedure

IT10

: Thermocouple,  J/

K/R/T, Selectable  ˚C/ ˚F,

           1˚ /0.1˚      

R

K/T

J

T/C +

J/K/R/T THERMOCOUPLE

T/C −

LINEARISATION IS DIGITAL

BL-40H
DL-40H
FL-B101D40H

LEOPARD

153A

Cold Junction Thermocouple 

Type Select Header

Summary of Contents for CL-B101D40-TCRTD

Page 1: ...or below Lo each setpoint Digital display blanking Decimal point setting Four level brightness control LEOPARD FAMILY Input Specs Depends on Input signal conditioner A D Converter 14 bit single slope...

Page 2: ...direction indicated by the arrow When an alternative dotted line is shown this indicates that an alternative logic branch will be followed when a particular option is present When two buttons are sho...

Page 3: ...999 seconds Pump PUM Hysteresis hYSt Hysteresis hYSt Hysteresis hYSt Setpoint 3 SP3 Setpoint 4 SP4 Operational Display MAIN MENU BRIGHTNESS AND SETPOINT VIEW MODE Bargraph Center Point Display Selecti...

Page 4: ...E STEP A Calibration Mode STEP B Calibration Mode To Digital Span Selection for Bargraph Display Page 5 cAL out iP branch will only appear if the analog output option is installed otherwise cAL on goe...

Page 5: ...between Cto and oFF the software is detecting that the optional analog output hardware is NOT installed Digital Span Selection For Bargraph Display Bargraph does not light up for Input Signals up to...

Page 6: ...igh and AnLo analog low can be reversed to provide a 20 to 4mA output The digital span selected can be as small as two counts when using the analog output to function as a Control or Alarm Driver Smal...

Page 7: ...nd 18 see Rear Panel Pinouts on page 10 Using the and buttons adjust the analog output to the desired low value as measured on the multimeter cLo may be adjusted to any value from 0 3 mA to 18 mA mA o...

Page 8: ...Press the P button Display toggles between PUM and on or oFF STEP F Select Pump PUM on or oFF 1 Using the and buttons select the Pump to be ON or OFF When PUM is selected ON and SP2 is set at a value...

Page 9: ...Bargraph Color when the bar is ABOVE SP2 Setpoint 1 Using the and buttons select the desired bargraph color grn oran or red 2 Press the P button Display toggles between CSP3 and the previous color se...

Page 10: ...p Form A relays An analog output module is also shown as installed The CL B101D40 TCRTD uses plug in type screw terminal con nectors for all input and output connections The power supply connections p...

Page 11: ...5A SP4 SP3 SP2 SP1 SP1 SP2 SP3 SP4 Relay Modules with 4 Isolated 5A Form A Relays Output Carrier Board Output Carrier Board Relay Module Relay Module Input Signal Conditioner Analog Output Module Mai...

Page 12: ...the back panel back into the case Apply power to the meter 4 Enter the program mode and select the type of thermocouple J K R T the resolution 0 1 or 1 and the display units C or F See Page 4 of the d...

Page 13: ...10 72VDC Input Modules Partial List See www texmate com IT10 Thermocouple J K R T Selectable C F 1 0 1 IT11 RTD 100 Pt Selectable 3 4 wire C F 1 0 1 385 392 IT15 RTD 1000 Pt Selectable 3 4 wire C F 1...

Reviews: