29
1. Activate external
trigger
Press the Shift key followed by the TRIG key. The
EXT indicator appears on the display.
2. Start trigger
Press the TRIG key to start triggering manually.
The
indicator turns On.
Reading indicator
The reading indicator
does not flash before triggering (can be on or off).
After triggering, the indicator flashes according to the external signal trigger
timing.
Exit external trigger
Press the Shift key followed by the TRIG key. The
EXT indicator disappears and the trigger goes
back to internal mode.
4-2-3. Set trigger delay
Background
Trigger delay defines the time rag between triggering and measurement start.
The default is set at 10ms.
Panel operation
1. Press the Shift key, the 2ND (Menu) key,
the Right key, the Down key. The delay
menu appears.
2.
Press the Down key. The delay setting
appears.
3.
Move the flashing point (cursor) using the
Left/Right key. Change the value using the
Up/Down key.
4.
Press the ENTER key to confirm editing
and press the EXIT key. The display goes
back to previous mode.
Range
1 ~ 1000ms, 1ms resolution