40
7-5. Command Set
Commands are
non
-case sensitive.
Underline means a single space (dc_1
→DC 1V).
When the parameter does not match the real value, the closest possible option is automatically selected
(dc_2 [DC 2V range]
→DC 10V)
7-5-1. CONFigure command
conf:volt:dc
Sets measurement to DC Voltage and specifies range.
Parameter: NR2, min, max
Example: conf:volt:dc_1 (DCV, 1V range)
Example: conf:volt:dc_min (DCV, minimum range)
conf:volt:ac
Sets measurement to AC Voltage and specifies range.
Parameter: NR2, min, max
Example: conf:volt:ac_1 (ACV, 1V range)
Example: conf:volt:ac_min (ACV, minimum range)
conf:volt:dcac
Sets measurement to DC+AC Voltage and specifies range.
Parameter: NR2, min, max
Example: conf:volt:dcac_1 (DC+ACV, 1V range)
Example: conf:volt:dcac_min (DC+ACV, minimum range)
conf:curr:dc
Sets measurement to DC Current and specifies range.
Parameter: NR2, min, max
Example: conf:curr:dc_10e-3 (DCI, 10mA range)
Example: conf:curr:dc_min (DCI, minimum range)
conf:curr:ac
Sets measurement to AC Current and specifies range.
Parameter: NR2, min, max
Example: conf:curr:ac_10e-2 (ACI, 100mA range)
Example: conf:curr:ac_min (ACI, minimum range)
conf:curr:dcac
Sets measurement to DC+AC Current and specifies range.
Parameter: NR2, min, max
Example: conf:curr:dcac_10 (DC+ACI, 10A range)
Example: conf:curr:dcac_min (DC+ACI, minimum range)
conf:res
Sets measurement to 2W Resistance and specifies range.
Parameter: NR2, min, max
Example: conf:res_10e3 (2W R, 10K range)
Example: conf:res_min (2W R, minimum range)
conf:fres
Sets measurement to 4W Resistance and specifies range.
Parameter: NR2, min, max
Example: conf:fres_10e3 (4W R, 10K range)
Example: conf:fres_min (4W R, minimum range)
conf:freq
Sets measurement to Frequency and specifies range.
conf:per
Sets measurement to Period and specifies range.
conf:cont
Sets measurement to Continuity.
conf:diod
Sets measurement to Diode.
conf:temp
Sets measurement to Temperature.
conf:stat:func?
Returns function of 1
st
display.
Parameter: 1 (DCV), 2 (ACV), 3 (DCA-10A), 4 (ACA-10A), 5 (DCA-mA), 6
(ACA-mA), 7 (2WR), 8 (Freq), 9 (TempC), 10 (AC+DCA-10A), 11 (AC+DCV), 12
(AC+DCA-mA), 13 (Diode), 14 (Period), 15 (TempF), 16 (4WR), 17 (Cont.)
conf:stat:rang?
Returns range of 1
st
display. Parameter:
DCV: 1 (100mV), 2 (1V), 3(10V), 4 (100V), 5 (1000V)
ACV: 1 (100mV), 2 (1V), 3(10V), 4(100V), 5(750V)
AC+DCV: 1 (100mV), 2 (1V), 3(10V), 4 (100V), 5 (1000V)
DCmA, ACmA, ACmA+DCmA: 1(10mA), 2(100mA), 3(1A)
2WR, 4WR: 1(100Ω), 2(1kΩ), 3(10kΩ), 4(100kΩ), 5(1MΩ), 6(10MΩ), 7(100MΩ)
DCA, ACA, AC+DCA (10A range): 1 (one range)
Freq, TempC, TempF, Diode, Period, Cont.: 1 (one range)
conf:auto
Set 1
st
display to Auto range.
Parameter: 0 (disable auto range), 1 (enable auto range)