91
3-14-34. :TRIGger:EXTERnal:PROBe:TYPe
Set
Query
Description
Sets or queries the external probe type.
Syntax
:TRIGger:EXTERnal:PROBe:TYPe { VOLTage |
CURRent | ? }
Related commands :TRIGger:EXTERnal:PROBe:RATio
Parameter
VOLTage
Voltage
CURRent
Current
Return parameter
Returns the probe type.
Example
:TRIGger:EXTERnal:PROBe:TYPe?
CURRENT
3-14-35. :TRIGger:EXTERnal:PROBe:RATio
Set
Query
Description
Sets or queries the external probe ratio (attenuation).
Syntax
:TRIGger:EXTERnal:PROBe:RATio {<NRf> | ?}
Related commands :TRIGger:EXTERnal:PROBe:TYPe
Parameter
<NRf>
External probe attenuation factor.
Return parameter
<NR3>
Returns the probe attenuation factor.
Example
:TRIGger:EXTERnal:PROBe:RATio?
5.01
3-14-36. :TRIGger:BUS:TYPe
Query
Description
Returns the current bus type.
Syntax
:TRIGger:BUS:TYPe?
Return parameter
I2C
I2C mode
SPI
SPI mode
UART
UART mode
PARALLEL Parallel mode
Example
:TRIGger:BUS:TYPe?
UART