81
3-14-6. :TRIGger:MODe
Set
Query
Description
Sets or queries the trigger mode.
Syntax
:TRIGger:MODe {AUTo | NORMal | ?}
Parameter
AUTo
Auto trigger (Untriggered roll)
NORMal
Normal trigger
Return parameter Returns the trigger mode.
Example
:TRIGger:MODe NORMal
Sets the trigger mode to normal.
3-14-7. :TRIGger:HOLDoff
Set
Query
Description
Sets or queries the holdoff time.
Syntax
:TRIGger:HOLDoff {<NRf> | ?}
Parameter
<NRf>
Holdoff time
Return parameter <NR3>
Returns the trigger holdoff time.
Example
:TRIGger:HOLDoff 1.00E-8
Sets the trigger holdoff time to 10ns.
3-14-8. :TRIGger:LEVel
Set
Query
Description
Sets or queries the level.
Syntax
:TRIGger:LEVel {TTL | ECL | SETTO50 | <NRf> | ?}
Related commands :TRIGger:TYPe
Parameter
<NRf>
Trigger level value
TTL
Sets the trigger level to TTL.
ECL
Sets the trigger level to ECL.
SETTO50 Sets the trigger level to the User level
(50% by default).
Return parameter
<NR3>
Returns the trigger level.
Example1
:TRIGger:LEVel TTL
Sets the trigger to TTL.
Example2
:TRIGger:LEVel 3.30E-1
Sets the trigger level to 330mV/mA.