Optional display for measured value display
46
Optional display for measured
value display
Description
A display module can also be plugged onto the control board for the
measured value display (see
on page 49).
The programming menu cannot be called up via the display. It is only
designed for the measured value display. Use the Service Monitor to
program the device. (see
on page 27).
Risk of damage to the display
Only plug the display onto the control board for long-term use. Re-
peated plugging and unplugging can cause damage to the contacts,
resulting in display failure. Disregard will result in the termination of
the warranty service.
Installation
➢
Switch the device off.
➢
Carefully plug the display into the designated slot on the main
board (see
Functions of the operating and display elements
on
page 20).
Make sure that all contacts click into place properly.
Notes on operation
When the device is switched on with the display plugged in, the in-
stalled software version is displayed during the boot process. This al-
lows you to visually check the current software version after a soft-
ware update.
After a measurement has been completed, the measurement result is
shown in the selected display unit. The display value is retained until
the end of the next measurement and is then updated to the new
measured value.
If a fault occurs during the last measurement, the display shows “x x
x x x selected display unit”. If it is a temporary fault, the Alarm func-
tion key flashes at the same time (see
on page 35).
In this case, the device will start a new measurement once the pro-
grammed analysis interval has expired.
If the error is still present, the indication on the display remains and
the Alarm function key remains illuminated. In this case, no further
measurements are made during programmed analysis interval and
the device switches to standby mode. If the fault is no longer present,
the measurement result is shown on the display once the measure-
ment has been completed and the device switches back to normal
operating mode.
flashing
illuminated
NOTE