Assembly
14
Remove the supplied funnel.
Accommodate the funnel
underneath, in the recess of the hous-
ing designated for that purpose.
Connect the funnel of the Testomat
®
Modul CL with a drain hose
(internal diameter 12 mm).
Lay this hose
free of back pressure
and without the siphon ef-
fect to the drain
.
Connect mains and devices
Connect the device only to a 24 VDC power supply.
Risk of injury from assembly when voltage is present!
Unless you disconnect the power supply before commencing assem-
bly, you risk injury, destruction of the product or damage to system
components.
Disconnect all power to the relevant system component before as-
sembling the Testomat
®
Modul CL device.
When connecting, use only verified cables with sufficient line
cross-section (see
Requirements of cable for operating voltage
and system components and installed lines
on page 6).
Danger of damage due to electromagnetic fields!
If you assemble the Testomat
®
Modul CL device or connecting cables
parallel to mains cables or in the vicinity of strong electromagnetic
fields, the device may be damaged or the measurement disrupted.
Keep the connecting cables as short as possible.
Lay the connecting cables and mains cables separately.
Shield the device from strong electromagnetic fields.
WARNING
!
CAUTION
!