6 Operating Instructions 

17

 

 

If the actual temperature display 18) shows 

, press 

 in the control 

field 21) during 2 sec to switch on the temperature controller. 

 

Press  

 and then 

 and/or 

 to change the temperature setpoint. If the desired 

value is indicated by the setpoint display 19), safe by pushing 

 again. 

 

Switch on the evaporation tube heating with toggle switch 17) to connect the controller 
power output to the heater. 

 

The green heating current OK LED 15) indicates that supply voltage (24 VDC) is applied 
to the heating and the current is within its tolerance. 

 

If one of the red heating current LED's 14) or 16) lights, the current is too low or too high, 
probably caused by a defective heating element or an interrupt in the electric circuit. The 
evaporation tube should be switched off immediately and the device has to be checked 
and repaired by the manufacturer or a local service provider. 

 

Wait until the measured gas temperature 18) agrees with the set value 19) /-

 

2

 

°C. 

During the heating up phase the heating current OK LED 15) lights continuously green. 
The LED starts blinking in intervals of a few seconds when the measured gas 
temperature approaches the set value. When the set value is reached the heating 
on/off-duty cycle stabilizes on a ratio depending on the adjusted temperature. 
Heating up times are approximately 2.5

 

min from ambient to 200

 

°C and 4

 

min to 300

 

°C. 

 

Due to attachment of low volatile substances on its inner surface, the evaporation 
tube may produce particles itself if operated at temperatures above the previous 
operation point. In this case, the attached volatile material evaporates, and may 
re-nucleate being measured by the connected instrumentation. When the 
evaporation tube temperature is increased, it is recommended to run the testo 
ASET15-1 with stopped rotating diluter disk but active testo MD19-3E dilution air 
supply to rinse the tube with filtered air until no undesired particles are detected 
anymore. This process may last up to 20 minutes. 

6.3.

 

Flows and Control LED's in Air Supply Part 

Note: All flows mentioned in this manual are standard volume flows in [l/min] which means liters 
per minute at standard conditions: 1013,25 hPa / 0°C. 
testo ASET15-1 in its pneumatic function is an adjustable diluter where two calibrated flows Q

MD

 

and Q

AS

 are mixed in the secondary dilution mixing chamber 37). Fig. 6.2 shall help to 

understand the setting of the secondary dilution air in the air supply part of testo ASET15-1. 

 

Fig.6.2: secondary dilution flows 

Measuring gas flow Q

MD

 (standard 1.5

 

l/min) from the testo MD19-3E Rotating Disk Diluter enters 

the secondary dilution mixing chamber. Secondary diluted measuring gas flow Q

MG

 is drawn from 

the instrumentation connected on the measuring gas output 13). 
It is evident that Q

AS

 must be adjusted to a value where Q

MD

 

+

 

Q

AS

 

>

 

Q

MG

 and therewith Q

EX

 

>

 

0 to 

ensure no gas is sucked backwards through the excess gas tube and added to the measuring 
gas. 
 

 

 

Summary of Contents for ASET15 1

Page 1: ...testo ASET15 1 User Manual...

Page 2: ...2...

Page 3: ...4 2 Definitions 9 4 3 Abbreviation Units and Symbols 9 4 4 The System 10 4 5 Control Elements and Connections 11 4 5 1 Important Remarks 11 4 5 2 Front View 11 4 5 3 Gas Connectors 12 4 5 4 Rear View...

Page 4: ...Control 22 8 Maintenance and Calibration 22 8 1 Evaporation Tube 22 8 2 Parts with Limited Lifetime 22 8 3 Storage Acclimatization 22 8 4 Operation Environment Requirements 23 9 Appendix 24 9 1 Exten...

Page 5: ...iginal manufacturer s warranty applies There are no user serviceable parts inside testo ASET15 1 and some very sensitive parts Do not open your testo ASET15 1 as you may damage it Warranty is voided i...

Page 6: ...esto SE Co KGaA 3 Safety 3 1 Risk Types The following diagram shows typical risks that could cause damage or injury while handling the testo ASET15 1 Air Supply Evaporation Tube Fig 3 1 risk types 3 1...

Page 7: ...about installation operation and maintenance Warning Warning means that improper operation could cause a serious human or instrument damage or injury with consequence of irrevocable instrument damage...

Page 8: ...The total measuring gas flow up to 16 5 l min enables the user furthermore to connect nanoparticle instrumentation which consumes higher measuring gas flows than can be drawn from the testo MD19 dilut...

Page 9: ...e outlet of the evaporation tube 4 3 Abbreviation Units and Symbols testo ASET15 1 Air Supply Evaporation Tube 15 15 l min air supply 1 version 1 CVS Constant Volume Sample fullstream dilution tunnel...

Page 10: ...evaporation tube inlet port 34 Thermally insulated evaporation tube 37 Secondary dilution mixing chamber 39 Heat sink The testo ASET15 1 is power supplied by one phase electricity An internal pump fee...

Page 11: ...emerging from other processes Electric Shock When in operation any electrical equipment can produce dangerous voltages Ignoring these warnings may result in serious injury or damage of the equipment...

Page 12: ...range 16 Heating current too low LED red current too low interrupt 17 Evaporation tube heating ON OFF swich 18 Actual evaporation tube temperature red 19 Evaporation tube temperature setpoint green 2...

Page 13: ...the CU 2 digital control unit 28 Fuse of the integrated testo MD19 3E 5A slow 29 Fuse of testo ASET15 1 5A slow 30 Mains switch 31 Mains connector 32 Additional evaporation tube inlet port closed Skin...

Page 14: ...1 Air Supply Evaporation Tube is a secondary dilution unit and cannot be operated without an inserted primary testo MD19 3E Rotating Disk Diluter 5 1 Integrating testo MD19 3E into the testo ASET15 1...

Page 15: ...f the plug cannot be inserted the fixation ring might be in the wrong position Push down the release button and insert the plug again The excess gas connection 12 meets the same standard but male and...

Page 16: ...of the testo MD19 3E pump might have an influence on the flow regulation in the Air Supply of the testo ASET15 1 The tube connected to testo MD19 3E pump outlet can fill with liquid from condensation...

Page 17: ...in to 300 C Due to attachment of low volatile substances on its inner surface the evaporation tube may produce particles itself if operated at temperatures above the previous operation point In this c...

Page 18: ...reading corresponds directly to the secondary dilution factor Examles E x a m p l e s pot setting 2 QAS 2 1 1 5 l min 1 5 l min dilution factor DF QAS QMD QMD DF 1 5 1 5 1 5 2 pot setting 5 QAS 5 1 1...

Page 19: ...rts to light to find the ideal measuring gas flow setting The secondary dilution can be used to increase the total dilution factor as well as to reduce thermophoretic losses If it is set as high as mo...

Page 20: ...contact of your socket If the plug is replaced ensure the yellow green ground wire of the cable is properly connected to the new ground pin or the case is otherwise connected to protective earth whic...

Page 21: ...remote 13 digital ground D 0 VDC for digital inputs outputs 14 secondary dilution air supply flow error LED 5 D O 0 VDC OK 5 VDC error 15 evaporation tube current Iheat too low LED 16 D O 0 VDC OK 5 V...

Page 22: ...r longer measuring programs 8 2 Parts with Limited Lifetime Filter media and moving parts like air supply pumps inside the testo ASET15 1 have limited lifetimes due to ambient air pollution and mechan...

Page 23: ...intended to be used outdoor or in a dusty or wet environment IP protection degree IP 20 testo ASET15 1 is protected against accidental contact to dangerous parts of the instrument It is not protected...

Page 24: ...nanoparticles inlet gas flow 1 5 l min actively fed to the primary diluter and returning from there accuracy 3 air supply flow 0 15 0 l min accuracy 3 of set value 0 1l min ex dilution factor 1 11 mea...

Page 25: ...61010 1 2001 Safety requirements for electrical equipment for measurement control and laboratory use and therefore is in conformity with the following European Directives in their current versions 20...

Page 26: ...0 001 1 3 937 10 8 1 1 in 0 0254 2 54 25 4 2 54 104 2 54 107 1 Temperature Celsius Fahrenheit C F 0 C 0 32 T C T F 32 1 8 100 C 100 212 T F T C 1 8 32 0 F 17 78 0 100 F 37 78 100 Mass Kilogram Gram Po...

Page 27: ...w 0 3 l min 12 Excess measuring gas output 13 Quick coupling for measuring gas output to sensor s 14 Heating current too high LED red current too high short circuit 15 Heating current OK LED green cur...

Page 28: ...ection between mixing chamber and heat sink 39 41 Fan for cooling the components integrated in the testo ASET15 1 42 Heat sink cooling fan 43 Outlet connection to front panel Fig front view of testo A...

Page 29: ......

Page 30: ...Testo SE Co KGaA Testo Strasse 1 79853 Lenzkirch Germany phone 49 7653 681 5062 fax 49 7653 681 95062 e mail sales nanoparticle testo de...

Reviews: