82
testo 6651 Ethernet
-
1.6 Maintenance and cleaning
1.6 Maintenance and cleaning
1.6.1 Maintaining the instrument
We recommend that the adjustment and settings of the transmitter be
checked at regular intervals using the
User menu (see
chapter 1.4, page 60
) or
P2A software (see
volume 2, chapter 3, page 113)
Convenient "remote monitoring" of the transmitter can be implemented, for
example by using a relay as a collective alarm (see
chapter 1.4.6.3, page 66
)
whose messages are forwarded to a local alarm transmitter (horn, light) or
PLC.
1.6.2 Cleaning the instrument
▪
Only clean the instrument carefully with a moist cloth.
▪
Do not use aggressive cleaning agents.
▪
Do not use any solvents.
▪
Do not touch or damage the sensor
1.888.610.7664
www.
Testo-Direct
.com