7 Maintaining the product
87
2.
Unlock the condensate trap / condensate container and pull it
vertically off the meas. box.
3.
Open the drain plug (1) and let the condensate run out into a
sink.
4.
Wipe off any drops still on the condensate outlet with a cloth
and close the condensate outlet.
5.
Plug the condensate trap / condensate container on the meas.
box.
7.10.
Checking / replacing the dirt filter
Checking the dirt filter:
> Check the dirt filter of the meas. box periodically for
contamination: Check visually by looking through the window of
the filter chambers. In case of visible contamination: Change
the dirt filter.
Replacing the dirt filter:
The filter chamber may contain condensate.