6 Using the product
75
3.
6.2.3.2.
Draught-Measurement
No flue gas probes with preliminary filter may be used for
draught measurements.
✓
A flue gas probe must be connected.
✓
The pressure socket of the instrument must be free
(depressurized, not closed).
Do not measure for longer than 5 min, as the drift of the
pressure sensor could have the effect that the readings are
outside the tolerance limits.
The
automatic pressure zeroing
(retrofit by Testo
Service) automatically zeroes the pressure sensor at
regular intervals (60s), to prevent the typical drift of the
pressure sensor.
>
[Options]
→
Save
: The readings are saved in a record.
>
[Options]
→
: The readings from a record are printed.
>
[Options]
→
Measurement view
: (This function is not available
during a measurement): The configure measurement view
menu is opened.
>
[Options]
→
Folders
: The folder
Folders
is opened.
>
[Options]
→
Show Graphic
: The readings are displayed in
form of a line graph.
>
[Options]
→
Configure Graphic
: The measurement
parameters to be represented (max. 4) can be displayed ( ) or
hidden ( ).
Calling up the function:
>
Measurement type
→
Draught-Measurement
→
[OK]
.
Performing the measurement:
1.
Start measurement:
[ ]
- Draught zeroing (7s).
- Rinse (approx. 10s).
2.
Position the flue gas probe in the hot spot (area of the highest
flue gas temperature). The display showing the maximum
measured flue gas temperature (FT) helps when positioning the
probe.
- The reading is displayed.
3.
Quit measurement
[ ]
.
- The reading is maintained.
Summary of Contents for 350 Maritime V2
Page 1: ...testo 350 Flue gas analyzer Instruction manual...
Page 2: ...2...
Page 106: ...0970 3511 en 05_GB...