Operation manual Pneumator
∙
Testo Industrial Services
∙
Version 1.4
∙
2019
∙
Page 13
Phase
number
Indication in
menu
Signification
1
Wait for start (activate with OK), equipment is venting
(display shows VENT)
2
T START
Waiting time (in seconds) for venting (display shows
VENT)
4
Zeroing phase (if AUTO ZERO = ON), about. 3-5
seconds. At AUTO ZERO = OFF
phase is dropped
7
T HOLD
Holding time (in seconds) at point zero
9
Adjustment of the next higher pressure stage
10
T HOLD
Holding time (in seconds) for adjusted pressure
(equivalent to the duration of phase 7)
13
T STOP
Waiting time (in seconds) at maximum pressure
(„range“)
15
Adjustment of the next lower pressure stage
16
T HOLD
Holding time (in seconds) at “pressure down” steps and
when reaching the zero point again (equivalent to the
duration of phase 7)
18
T PAUSE
Pause time between 2 cycles
Operation / application (AUTO):
After selecting the operation mode AUTO and leaving the menu (MENU button) the
AUTO display appears which supplies all relevant information for the automated
calibration procedure. The device is now in phase 1
(Waiting for OK = start).
The meaning of the display elements (AUTO):
o
Top left: Indication of the current operation mode
o
VENT = venting
o
ZERO = zeroing
o
CTRL = Adjustment and holding of the programmed pressure stages