3.6 AOT processing
3.6.1 Principle
The processing of AOT data retrieves old measurements and re-calculate the
optical thickness based on new calibration coefficients.
This will usually happen after a new calibration of the photometer. It can also be
carried out on an experimental basis.
3.6.2 Re-calculation with software
After loading the data you want to work with, into the computer, click the tool icon
AOT Processing
.
The software displays the original AOT curves (dotted) and the curves recalculated
from current coefficients in solid lines.
Here's what can be achieved on the screen :
3.6.3 Change settings
The role of the parameters is quickly and easily readable: the curves are redrawn
every variation of parameters.
CN
0
with arrows
The coefficients of Rayleigh and ozone only with an editor.
Visualization of curves
However, for a re-calculation of AOT, you do not have a setting to change and they
have already been updated with the new calibration.
Selected AOT
initial curves of
the day
(level 1.0 ou 1.5)
Dotted lines.
Recalculated AOT
curves in solid
lines.
Parameters
changed by user
Recalculated AOT values
( NAOT465, NAOT540 et NAOT619 )
Initial AOT values
( AOT465, AOT540 et AOT619 )